Tag Archives: Calibration

A solid 2.5 ns reference for FLIM-phasors calibration

We measured the solid PMMA reference slide from the FLIM-Phasors Calibration Kit against fluorescein. It has a monoexponential lifetime of 2.49 ns. Used as the calibration reference, it gives the same lifetime maps and phasor populations in Convallaria as fluorescein does, so you can calibrate without preparing a dye solution.