Home › Methods › Geometric (1/e) Method

Methods

Geometric (1/e) Method

The 1/e method is a fast, model-free and geometric approach: the lifetime is read directly from the decay curve as the time at which the intensity drops to 1/e of its peak value.

For a mono-exponential decay this time equals τ, while for multi-exponential decays it gives an effective lifetime. It requires no fitting, but depends on a well-defined peak and a good signal-to-noise ratio around the 1/e point.

Geometric (1/e) method

Figure 1 · The lifetime is the time at which the decay falls to 1/e (about 37%) of its peak. For a mono-exponential decay this time equals τ; for a multi-exponential decay it gives an effective lifetime, τeff. Schematic.

From principle to measurement

One point, no fitting

Reading the lifetime at 1/e needs no model, only a clean decay with enough photons. FLIM LABS TCSPC acquisition records the photon-resolved decays this approach reads, from a cuvette or from every pixel.